HP 8360 User Manual

Page 107

Advertising
background image

210

OUTPUT

220

ENTER

230

PRINT Id$

240

PRINT

250 !
260

PRINT "Counter Used . . .

270

OUTPUT

280

ENTER

290

PRINT Id$

300

PRINT

310 !
320

OUTPUT

ON"

330 !
340

PRINT

350

PRINT "INPUT

360

____________

370

PRINT

380 !
390

FOR Testpoint=First TO Last

400

OUTPUT

410

ENTER

420

OUTPUT

430

ENTER

440

PRINT Testpoint,Reading/lOOO

450

NEXT Testpoint

460 !
470

OUTPUT

OFF"

480

END

Program Comments. Lines 20 to 70: Declare variables and I/O paths
for instruments. I/O paths let you use a name for an instrument in
OUTPUT and ENTER

of a numeric address.

80 to 100: Assign values to the input test limits in

110 to 130: Clear the instrument HP-IB interfaces.

140 to 160: Reset each instrument to a known measurement state.

170 to 190: Print the test report title.

200 to 310: Query measurement instruments’ identifications for test
traceability.

320 to 330: Connect the source output signal to the output
terminals.

340 to 380: Print results table header.

390 to 460: This is the main measurement loop. Line 400 contains
two commands. :SOURce:VOLT sets the output level of the source.

is used to signal that the preceding command has finished

executing. To make an accurate

source output

must be allowed to settle. When the output has settled, *OPC? places

Getting Started Programming 1-77