Entry on/off, Ext det cal – HP 8360 User Manual

Page 202

Advertising
background image

Ext Det Cal

Function Group

Menu Map

Description

Programming Codes

See Also

ENTRY

NONE

This

lets you turn off (blank) the active entry area and

disable the ARROW keys, rotary knob, and entry keys. When
any function key (hard or soft) is pressed, the active entry area is

reactivated. The yellow LED, ENTRY ON, next to

indicates whether the entry area is active (LED on=active).

SCPI:

No specific code activates

Arrow Keys

“Entry Area” in Chapter 1.

Ext Det Cal

Function Group

USER CAL

Menu Map

Description

This

enables the synthesizer to act as a controller to an HP

power meter. This

causes an immediate execute on

the interface bus and generates an HP-IB error if no power meter is
present on the interface bus or if the synthesizer is unable to address
the power meter. Use external detector calibration to characterize
and compensate for an external negative diode detector used in an
external leveling configuration.

Programming Codes

CALibration:PMETer:DETector:INITiate?
CALibration:PMETer:DETector:NEXT?

suffix]

See Also

“Optimizing Synthesizer Performance” in Chapter 1.

HP 8360
User’s Handbook

Operating and Programming Reference