Internal Pulse Generator Period

Internal Pulse Mode Gate

Turns on the internal pulse mode during the
positive cycle of the externally generated pulse.

Internal

Pulse Mode Trigger

Triggers on the leading edge of the external pulse
input.

Programming Codes

SCPI:

NONE, see the individual

listed.

Analyzer:

NONE

See Also

a so see “Modulation” and “Pulse”.

Internal Pulse Generator Period

Function Group

Menu Map

Description

This

(Option 002 only) lets you set a value for the internal

pulse generator’s pulse period. The pulse is adjustable from
300 ns to 400 ms with ‘25 ns resolution. The factory preset default
is 2 ms pulse period. When this feature is active, its current value is
displayed in the active entry area.

Since period and rate are inversely related, if both are given values,

only the last one will be applied which will cause the first one to be
recalculated. Use the one that is convenient for your application. For
example, if you set the pulse period, do not change the pulse rate

(the synthesizer automatically adjusts the rate to match the period.)

Programming Codes

SCPI:

PULS:INTernal:PERiod <num>[time

Analyzer:

NONE

See Also

a so see “Pulse” and “Modulation”.

HP 8360
User’s Handbook

Operating and Programming Reference