HP 8360 User Manual

Page 65

Advertising
background image

Setup Synthesizer Parameters

6.

On the synthesizer, press (PRESET).

7.

FREQUENCY

8.

Access User Flatness Correction Menu

9.

10.

11.

12

13

14.

Press POWER

Select

Menu.

Select

Delete

Menu

Delete All

. This step insures that the

flatness array is empty.

Press

Leave the delete menu and return to the previous

soft key menu.

Enter the frequency points at which the correction information
will be taken. Choose either the point-by-point entry method

Enter Freq or the automatic frequency point generation

Auto Fill Start.

For this example,select

Auto Fill Start

Select Auto Fill Stop

Auto Fill

Notice that a frequency list starting at 4 and ending at 10
with an increment value of 1

is created.

Enter Correction Data into Array

Select

Menu Measure

All.

The power meter

is now under synthesizer control and is performing the sequence
of steps necessary to generate the correction information at each
frequency point.

If an HP-IB error message is displayed verify that the interface
connections are correct. Check the HP-IB address of the power
meter and ensure that it is the same address the synthesizer is
using (address 13 is assumed). Refer to the menu map 8, System,

for the key sequence necessary to reach

Meter Adrs .

Enable User Flatness Correction

15.

When the operation is complete, (a message is displayed) the
flatness correction array is ready to be applied to your setup.
Disconnect the power meter/sensor and press

(amber LED on). The power produced at the point where the
power meter/sensor was disconnected is now calibrated at the

frequencies and power level specified above.

Getting Started Advanced

Advertising