HP 8360 User Manual

Page 67

Advertising
background image

Figure 1-19. Creating a User Flatness Array

For this example, refer to menu map 5, POWER.

1.

2.

3.

4.

5.

6.

7.

8.

The equipment setup shown in Figure 1-19 assumes that if your
setup has an external leveling configuration, the steps necessary
to correctly level have been followed. If you have questions about
external leveling refer to earlier paragraphs titled, “Externally
Leveling the Synthesizer

Setup

Power Meter

Zero and calibrate the power meter/sensor.

Connect the power sensor to the point where flatness corrected
power is desired.

Setup Synthesizer Parameters

On the synthesizer, press (PRESET).

(

P O W E R L E V E L

)

This sets the test port power to

Create A Frequency List

On the synthesizer, press FREQUENCY

Select List Menu Enter

This enters 5

as the first frequency in the list array. Entering a frequency

automatically sets the offset to 0

and the dwell to 10 ms.

Enter 18, 13, 11, and 20

to complete this example array.

Getting Started Advanced 1-37