HP 8360 User Manual

Page 330

Advertising
background image

l

DIAGnostics:INSTrument:PRINter:ADDRess

l

DIAGnostics:INSTrument:PRINter:ADDRess?

Sets the HP-IB address of the printer to use during some of the
calibration procedures when the synthesizer assumes HP-IB control.
*RST and power on do not effect this command. The default is 1.

The default value is set at memory initialization only.

l

DIAGnostics:IORW

Performs a write to the I/O Device number specified in the first

<num> and sets it to the value in the second <num>.

l

DIAGnostics:IORW?

Reads from the specified I/O device number and returns the response
data.

l

DIAGnostics:OUTPut:FAULt?

Returns a string of 16, and that are equivalent to the fault
display.

Bit 0 = PEAK
Bit 1 TRACK
Bit 2 = RAMP
Bit 3 = SPAN
Bit 4 =

Bit 5 = ADC
Bit 6 = EEROM
Bit 7 PWRON
Bit 8 = CALCO
Bit 9 PLLZERO
Bit 10 PLLWAIT
Bit 11 = FNXFER
Bit 12 = CAL YO
Bit 13 CAL MAN
Bit 14 = TMR CNFLCT
Bit 15 SEARCH

l

DIAGnostics:RESult?

Returns the following information:

<result>, <test failure>, <manual entry point>

where,

is one of the following:

0

Diagnosis successful.

1

Cannot diagnose; full

must be executed first.

2

No failures found-all selftests passed.

- 1

Cannot diagnose; diagnosis routine failed to isolate failure.
(software fault)

<test failure> is the test number of the most relevant failure
(-999 if parameter is not used, as in <result> of 1).

Operating and Programming Reference

HP 8360

User’s Handbook

Advertising