Info 2

Programming Codes

SCPI:

NONE

Analyzer:

NONE

See Also

Fault Menu

Fault Info 3

Function Group

SERVICE

Menu Map

Description

This

displays the latched status of the following fault

messages.

CALYO FAIL

Indicates that the YO adjusted at power-on or at
preset is unable to calibrate. Initiate a full self-test
to gather more information if this fault is indicated.

Indicates that the manual sweep DAC adjusted at
power-on or at preset is unable to calibrate. Initiate
a full self-test to gather more information if this fault
is indicated.

FAIL

T M R
CNFLCT
FAIL

SEARCH

Indicates a possible internal software error. Two
routines are trying to use the same timer.

Indicates that the ALC search leveling algorithm has
failed. This fault indication is possible only if the
search leveling mode is on.

Programming Codes