HP 8360 User Manual

Page 73

Advertising
background image

Note

Scalar Analysis Measurement with User Flatness Corrections,
Example 4

The following example demonstrates how to setup a scalar analysis
measurement (using an HP 8757 Scalar Network Analyzer) of a 2 to
20

test device such as, an amplifier. User flatness correction

is used to compensate for power variations at the test port of a

directional bridge. Follow the instructions to set up the synthesizer,
then configure the system as shown in Figure 1-21.

The synthesizer’s rear panel language and address switches must be
set to 7 and 31 (all l’s), to change the language or address of the
synthesizer from the front panel. The programming language must be
set to Analyzer. Refer to menu map 8, System, to find the location

of

Programming Language Analyzer

(asterisk on = active

language).

SYNTHESIZER

DETECTOR

DETECTOR

OIRECTIONRL

DETECTOR

I

Figure

1. Scalar System Configuration

Example Overview

In this example you use an HP

power meter to automatically

enter correction data into the array. It is necessary to turn off
the HP 8757 System Interface (controlled from the front-panel of
the analyzer) so that the synthesizer can temporarily control the
power meter over HP-IB . When the correction data entry process
is complete, enable user flatness correction and set the desired test
port power level. Then store the correction table and synthesizer
configuration in the same register that contains the analyzer
configuration. Re-activate the HP 8757 System Interface and recall

Getting Started Advanced 1-43

Advertising