CAUTION

Before switching on this instrument, make sure that the line

voltage selector switch is set to the voltage of the power supply and
the correct fuse is installed.

n

Always use the three-prong ac power cord supplied with this

instrument. Failure to ensure adequate earth grounding by not
using this cord may cause instrument damage.

n

Before switching on this product, make sure that the line voltage

selector switch is set to the voltage of the power supply and
the correct fuse is installed. Assure the supply voltage is in the
specified range.

Ventilation Requirements:

When installing the instrument in a

cabinet, the convection into and out of the instrument must not be
restricted. The ambient temperature (outside the cabinet) must be

less than the maximum operating temperature of the instrument

by 4

for every 100 watts dissipated in the cabinet. If the total

power dissipated in the cabinet is greater than 800 watts, then
forced convection must be used.

n

This product is designed for use in Installation Category II and

Pollution Degree 2 per IEC 1010 and 664, respectively.

Note

The detachable power cord is the instrument disconnecting device.

It disconnects the mains circuits from the mains supply before other
parts of the instrument. The front panel switch is only a standby
switch and is not a LINE switch.

vi