POWER LEVEL

),

the ALC level and attenuator are set automatically to

provide the most accuracy for the power requested.

Uncoupled Operation.

In some applications it is advantageous to

control the ALC level and attenuator separately, using combinations

of settings that are not available in coupled operation. In uncoupled
mode

when the desired power output is set via

(

POWER LEVEL

),

only the ALC level is changed. The attenuator setting

is changed via Set:

One use of uncoupled operation is power sweep, where the output

power linearly tracks the sweep voltage ramp. The synthesizer can
generate power sweeps of up to 40

depending on frequency. The

power at the start of the sweep is set via

(

POWER LEVEL

)

(coupled

operation) or by a combination of

POWER LEVEL

and Set

(uncoupled operation). The sweep range is entered by selecting

Power Sweep. If the sweep range entered exceeds the ALC range

(stop power greater than maximum available power), the UNLVLED

warning message appears at the end of sweep. No warning is given at
the time of entry. If the start power is entered when the synthesizer

is in coupled operation, the ALC level is set no lower than -10
limiting the available power sweep range. Using uncoupled operation
and setting the ALC level to -20

gives an additional 10

of

sweep range.

External Leveling

Leveling Mode

Normal ,

Leveling Point

or

or

Module

In externally leveled operations, the output power from the
synthesizer is detected by an external sensor. The output of this
detector is returned to the leveling circuits, and the output power is
automatically adjusted to keep the power constant at the point of
detection. Figure

shows a basic external leveling arrangement.

The output of the detected arm of the splitter or coupler is held
constant. If the splitter response is flat, the output of the other
arm is also constant. This arrangement offers superior flatness over

internal leveling, especially if long cables are involved. Flatness

may be improved with user flatness correction

((

FLTNESS ON

/

OFF

),

Menu applied at the external leveling point.