Step Swp Pt
Trig Ext

Function Group

Menu Map

Description

Programming Codes

See Also

FREQUENCY

2

When this

is selected, the synthesizer steps to the next point

in the stepped frequency sweep when an external hardware trigger is
received. When the last frequency point is reached and continuous
sweep is selected, the next trigger causes the step sweep to return to
the start frequency. Connect the trigger signal to the TRIGGER
INPUT BNC. An asterisk next to the key label indicates that this
feature is active.

SCPI: SWEep:TRIGger:SOURce EXT
Analyzer: TS

Step Swp Menu, Sweep Mode Step

“Using Step Sweep,” in Chapter 1.

Function Group

Menu Map

Description

Programming Codes

FREQUENCY

NONE

This

activates swept frequency mode and makes the stop

frequency parameter the active function. The start/stop frequency
must be separated by at least 2 Hz in order to remain in the
frequency sweep mode. If start=stop frequency then the zero span
mode is entered.

FREQuency:STOP <num>[freq suffix] or

FREQuency:MODE:SWEep

Analyzer: FB <num>

HP 8360