Agilent Technologies Option H48 Multiport Test Set Z5623A User Manual

Page 102

Advertising
background image

5 -34

Agilent Technologies Z5623A Option H48

Performance Verification

Setting Up Limit Testing

Figure 5-37

The Scale Window With Desired Values For Crosstalk

Setting Up the Limit Table

Step 1. Click Trace on the menu bar. When the Trace menu appears, scroll

down to the Limit Test... button and click it. See

Figure 5-8

.

Step 2. When the Limit Test window pops up as shown in

Figure 5-9

, click the

Show Table button, then click the OK button.

Step 3. When the Limit Table pops up as shown in

Figure 5-10

, transfer entries

from

Table 5-7

to the Limit Table on the network analyzer. (Consult the

shortcuts listed on

page 5-13

.) The result is shown in

Figure 5-38

.

Figure 5-38

Completed Limit Table for Crosstalk

Table 5-7

Limit Table Entries for Crosstalk

Segment

Number

Segment

Type

Beginning

Stimulus

Ending

Stimulus

Beginning

Response

Ending

Response

1

Max

300 kHz

1.3 GHz

−115 dB

−115 dB

2

Max

1.3 GHz

3 GHz

−110 dB

−110 dB

3

Max

3 GHz

6 GHz

−100 dB

−100 dB

4

Max

6 GHz

9 GHz

−97 dB

−97 dB

Advertising