Agilent Technologies Option H48 Multiport Test Set Z5623A User Manual

Page 90

Advertising
background image

5 -22

Agilent Technologies Z5623A Option H48

Performance Verification

Setting Up Limit Testing

Saving the Limit Test File

Step 1. Click File on the menu bar. When the File menu appears, scroll down to

the Save As... button and click it. See

Figure 5-14

.

Step 2. When the Save As window comes up as shown in

Figure 5-15

, type a file

name that includes the words, “return loss,” the words, “port active”,
and the date of the calibration listed in

Table 4-1 on page 4-13

. For

example, if the calibration file is named, “cal_06_17_01”, an appropriate
file name for this step would be, “return_loss_port_active_with_
06_17_01_cal”. See

Figure 5-16

. After entering the file name, record it

in

Table 5-3 on page 5-7

. Next, click the Save button.

Checking the Limit Test File

Step 1. Connect an RF cable from Port 1 of the network analyzer to Test Port 1

of the test set. Connect an RF cable from Port 2 of the network analyzer
to the Reflection Port of the test set.

Step 2. If the test set is turned off, turn it on. If the test set is already turned

on, turn it off, then back on again. This causes the test set to execute
the “*rst;” command, which is equivalent to the “refl_01;” command
combined with the “tran_08;” command.

Step 3. Recall the calibration file listed in

Table 4-1 on page 4-13

. (Consult

How to Recall a Calibration

” on

page 4-14

.) The resulting display

should look something like

Figure 5-17

.

Advertising