Environmental preparations – Agilent Technologies Option H48 Multiport Test Set Z5623A User Manual
Page 23
Agilent Technologies Z5623A Option H48
2 -7
Installation
Environmental Preparations
Environmental Preparations
1. Ensure that the
“Environmental Requirements” on page 1-6
are met.
2. If you are installing the test set into a cabinet, ensure there are at
least two inches of clearance around the sides and back of the test
set and the system cabinet. See
Figure 2-3
. The convection into and
out of the test set must not be restricted. The ambient temperature
(outside the cabinet) must be less than the maximum operating
temperature of the test set by 4
°C for every 100 watts dissipated in
the cabinet.
Figure 2-3
Ventilation Clearance Requirements
CAUTION
If the total power dissipated in the cabinet is greater than 800 watts,
forced convection must be used.