Agilent Technologies Option H48 Multiport Test Set Z5623A User Manual

Page 56

Advertising
background image

4-6

Agilent Technologies Z5623A Option H48

Calibrating the Network Analyzer

Performing the Calibration

Or repeat your cal.”

Figure 4-10

Cal Wizard Window After Cal Is Complete

Step 14. Get a pen or pencil and a piece of paper.

Step 16

will ask you to write

down a filename.

Step 15. Click the OK button. The calibration will now be saved to disk in the

following directory: C:\Program Files\Agilent\Network
Analyzer\Documents. The file will have a “.cst” suffix.

Step 16. Copy down the name of the file into which the calibration is saved. The

name will appear on the screen for about 10 seconds. Copy the filename,
not the path. (The path is listed in

Step 15

above.) A typical filename

would be, “at006.cst”. Record also the date and time that the calibration
was performed. (The date and time can be used to locate the file in case
the filename you write down is incorrect or incomplete.)