Performance Verification

Setting Test Limits

Applying

Equation 5-1

to the uncertainties listed in

Table 5-1

and to the

guaranteed specifications listed in

Table 1-1 on page 1-4

, we obtain the

test limits in

Table 5-2

:

Table 5-2

Derivation of Test Limits for Performance Verification of Test
Set

Parameter

Specification

Measurement
Uncertainty

Test Limit

Crosstalk

a,b

1) –110 dB

c

2) –105 dB

d

3) –95 dB

e

4) –90 dB

f

1) 5 dB

c

2) 5 dB

d

3) 5 dB

e

4) 7 dB

f

1) –115 dB

c

2) –110 dB

d

3) –100 dB

e

4) –97 dB

f

Return Loss

(port active)

g

1) 24 dB

c

2) 18 dB

d

3) 14 dB

e

4) 9 dB

f

1) 1.5 dB

c

2) 1.0 dB

d

3) 0.5 dB

e

4) 0.5 dB

f

1) 25.5 dB

c

2) 19 dB

d

3) 14.5 dB

e

4) 9.5 dB

f

Return Loss

(port off)

h

1) 26 dB

c

2) 20 dB

d

3) 16 dB

e

4) 14 dB

f

1) 1.5 dB

c

2) 1.5 dB

d

3) 1.0 dB

e

4) 0.5 dB

f

1) 27.5 dB

c

2) 21.5 dB

d

3) 17 dB

e

4) 14.5 dB

f

Insertion

Loss

i

1) 1.5 dB

c

2) 2.0 dB

d

3) 2.5 dB

e

4) 3.5 dB

f

0.3 dB
(all bands)

1) 1.8 dB

c

2) 2.3 dB

d

3) 2.8 dB

e

4) 3.8 dB

f

a. Between any two non-connected signal paths

b. Conditions: IF Bandwidth = 100 Hertz, Averaging = 10
c. Band 1 (300 kHz to 1.3 GHz)
d. Band 2 (1.3 GHz to 3.0 GHz)
e. Band 3 (3.0 GHz to 6.0 GHz)
f. Band 4 (6.0 GHz to 9.0 GHz)
g. When using an external termination on the Reflection or

Transmission port

h.When using one of the 50

Ω internal terminations

i. From any test set port to the Reflection or Transmission port

Advertising