Suggested external circuitry – Agilent Technologies Option H48 Multiport Test Set Z5623A User Manual

Page 119

Advertising
background image

Agilent Technologies Z5623A Option H48

7 -7

Advanced Topics

Using the Control Lines Connector

Figure 7-1

Control Lines

Suggested External Circuitry

Figure 7-2

shows one possible set of external circuitry for use with the

Control Lines connector. The integrated circuit (IC) shown in the figure
is a quad single-pole-single-throw (SPST) CMOS analog switch
(Temic/Siliconix Part No. DG413). The IC contains two normally open
and two normally closed switches.

Notice that, when “C1_in” is in the “Open” state, there will be no
current through the 10 k

Ω resistor and, consequently, no voltage drop

across it. This means that +15 V will be applied to the control line when
it is in the “Open” state. (The maximum allowable is +22 V.) When
“C1_in” is in the “Low” state, it will sink (15 V - 0.63 V) / 10 k

Ω = 1.44

mA of current. (The maximum allowable is 250 mA per control line.)
The values computed in this paragraph do not depend on V

C

, the

voltage used to control the DUT. (V

C

is applied to pin 3 in

Figure 7-2

.)

1

5

9

6

C1

C2

C3

C4

C5

Ground

Advertising