4-4

Agilent Technologies Z5623A Option H48

Calibrating the Network Analyzer

Performing the Calibration

Figure 4-4

Cal Type Window When It Is First Opened

Step 6. Click the Full SOLT 2-Port button. The result is shown in

Figure 4-5

.

Figure 4-5

Cal Type Window After Selecting “Full SOLT 2-Port”

Step 7. Click the OK button. The Cal Type window will then close and the Cal

Wizard window will pop back up as shown in

Figure 4-6

.

Figure 4-6

Cal Wizard Menu After Selection Of Cal Type Is Complete

Step 8. Verify that the currently selected calibration type shown in the Cal

Wizard window is “Full SOLT 2-Port (Omit Isolation)” as shown in

Figure 4-6

.

Step 9. Click the Measure ECal Stds button in the Cal Wizard window. The

window shown in

Figure 4-7

will then pop up.