Agilent Technologies Z5623A Option H48

5 -5

Performance Verification

Test Strategy

Insertion Loss

For this measurement, Port 1 of the network analyzer will be connected
to one of the test ports of the test set, and Port 2 of the network
analyzer will be connected either to the Reflection port or to the
Transmission port of the test set. By definition,

Insertion Loss =

− S

21

where Insertion Loss and S

21

are both in units of dB.

The network analyzer will be configured to measure S

21

. Limit lines

will be set up in such a way that each limit equals the negative of the
corresponding insertion loss test limit shown in

Table 5-2

. For example,

if an insertion loss test limit happens to be 1.0 dB, the corresponding
S

21

limit will be

−1.0 dB.

Insertion loss is a maximum spec. Therefore, each S

21

limit will be set

up as a minimum limit. For example, if an insertion loss test limit
happens to be 1.0 dB, the limit table will be set up so that S

21

must be

above

−1.0 dB in order for the instrument to pass. (S

21

will always be

zero or negative because the test set provides no amplification. In units
of dB, S

21

can range from zero to minus infinity.)

Return Loss

For this measurement, Port 1 of the network analyzer will be connected
to one of the test ports of the test set, and Port 2 of the network
analyzer will be connected to the Reflection port of the test set. By
definition,

Return Loss =

− S

11

where Return Loss and S

11

are both in units of dB.

The network analyzer will be configured to measure S

11

. Limit lines

will be set up in such a way that each limit equals the negative of the
corresponding return loss test limit shown in

Table 5-2

. For example, if

a return loss test limit happens to be 20 dB, the corresponding S

11

limit

will be

−20 dB.

Return loss is a minimum spec. Therefore, each S

11

limit will be set up

as a maximum limit. For example, if a return loss test limit happens to
be 20 dB, the limit table will be set up so that S

11

must be below

−20 dB

in order for the instrument to pass. (S

11

will always be zero or negative

because the test set provides no amplification. In units of dB, S

11

can

range from zero to minus infinity.)