Agilent Technologies Z5623A Option H48

6 -3

Measuring Multiport Devices

Calibrating the Test System

Figure 6-1 Calibrating the Test System

For the best accuracy, you should perform a full two-port calibration
between ports 1 and 3 on the test set, and again between ports 1 and 5.
As mentioned before, you need to save the calibrations as instrument
states. For information on how to calibrate and store instrument states,
refer to the PNA Series Network Analyzers User Guide, part number
E8356-90001.

CAUTION

Do not use the test set to change the RF signal path direction when you
are using a full two-port calibration. Doing so will render the
calibration invalid. Instead, use the internal transfer switch in the

analyzer.

GPIB

E8358A

(300 kHz - 9.0 GHz)

Port 1

Port 2

1

2

3

4

5

6

Reflection

Transmission

Calibration and Test

7

8

GPIB

A

B

Input

Calibration Paths