Agilent Technologies Option H48 Multiport Test Set Z5623A User Manual

Page 50

Advertising
background image

3-18

Agilent Technologies Z5623A Option H48

Using the Network Analyzer to Control the Test Set

Disaster Recovery

Figure 3-15

GPIB Device Disabling Message

Step 10. Right-click on the “AT-GPIB/TNT” icon once again. A menu will pop up.

Step 11. Click Enable on the menu.

Step 12. Close the Device Manager window by clicking on the X in the upper

right-hand corner of the window.

Step 13. Close the System Properties window by clicking on the X in the upper

right-hand corner of the window.

The recovery procedure is now complete. To resume control of the test
set, perform the procedure described in

“Preparing the Analyzer to

Control the Test Set” on page 3-2

.

Malfunctioning Network Analyzer

Symptoms

Any abnormality in the network analyzer’s behavior indicates a
malfunction of the analyzer. An example would be the failure of icons to
appear on the screen of the analyzer.

Recovery Procedure

Perform a Hard Reset on the PNA Series network analyzer as follows:
First, make sure the power is on (indicated by the power switch being
green). Next, hold the power switch in for about 20 seconds, or until the
color of the switch changes from green to orange. Release the switch,
then press it again and release it immediately.

Advertising