Agilent Technologies Z5623A Option H48

4-5

Calibrating the Network Analyzer

Performing the Calibration

Figure 4-7

Electronic Calibration Window

Step 10. Click the Module button. The window shown in

Figure 4-8

will then

pop up.

Figure 4-8

Module Connection Message

Step 11. Click the OK button. (The connections requested in

Figure 4-8

were

made in

Step 1

of “

Performing the Calibration

” on

page 4-2

.) This

window will close, the window shown in

Figure 4-7

will pop back up,

and the calibration will begin.

The network analyzer will complete the calibration with no further
operator actions required. The calibration will take about 5 seconds.

Step 12. Verify that the calibration has been completed. This is shown by a check

mark over the Module button as shown in

Figure 4-9

.

Figure 4-9

Electronic Calibration Window After Cal Is Complete

Step 13. Click the OK button shown in

Figure 4-9

. The window will close; the

Cal Wizard window will pop back up with the following message as
shown in

Figure 4-10

: “Done. Save your settings now by pressing OK”.