Agilent Technologies Z5623A Option H48

4-3

Calibrating the Network Analyzer

Performing the Calibration

Step 2. Press the Preset button on the Network Analyzer.

Step 3. Verify that the Ready light on the ECal Module is green.

Step 4. On the network analyzer, click Calibration on the menu bar, then click

the Calibration Wizard button. See

Figure 4-2

.

Figure 4-2

Calibration Menu on Network Analyzer

Step 5. When the window shown in

Figure 4-3

comes up, click the Cal Type

button. The Cal Type window shown in

Figure 4-4

will then pop up.

Figure 4-3

Cal Wizard Window

The currently selected calibration type is “None”.