Performing the calibration – Agilent Technologies Option H48 Multiport Test Set Z5623A User Manual

Page 52

Advertising
background image

4-2

Agilent Technologies Z5623A Option H48

Calibrating the Network Analyzer

Performing the Calibration

Performing the Calibration

Step 1. Make the connections shown in

Figure 4-1

.

The figure shows two RF cables connecting the two ports of the network
analyzer to the two ports of the ECal Module. The 24” RF cables
(Agilent Part No. 8120-4781 or equivalent) listed in

Table 2-2

on

page 2-3

should be used. Make sure the connectors are snugly tightened

but not overtorqued. A torque wrench set to 12 inch-pounds (Agilent
Part No. 8710-1766 or equivalent) is recommended.

Figure 4-1

Calibrating the Network Analyzer

Network Analyzer

Port 2

RF Cables

USB Port

Port 1

USB Quad Hub

Port 1 Port 2 Port 3 Port 4

Up Link

Keyboard with USB Cable

Mouse with

USB Cable

To USB Port

on Rear Panel

USB Cable

Electronic

Calibration

Module

USB Cable