5 -13

Performance Verification

Setting Up Limit Testing

Shortcuts:

a. To specify a Segment Type, click the Segment Type box in the

desired row, then use the drop-down menu.

b. To specify a Beginning Stimulus or an Ending Stimulus, click the

respective box in the desired row, then enter the desired value
followed by “k” if the value is in kilohertz, or by “g” if the value is in
gigahertz. For example, to specify a Beginning Stimulus of 300 kHz,
type, “300k” (without the quotation marks). To specify an Ending
Stimulus of 1.3 GHz, type, “1.3g” (without the quotation marks).

c. To specify a Beginning Response or an Ending Response, click the

respective box in the desired row, then enter the desired value in dB,
then hit the space bar. For example, to specify a Beginning Response
of

−1.8 dB, type, “−1.8<space>”.

d. Use the tab key to advance from left to right across a row. Use

<shift><tab> to advance from right to left across a row. (When
moving from right to left, the cursor will not move until you release
the shift key.)

Figure 5-10

Empty Limit Table

Figure 5-11

Completed Limit Table for Insertion Loss

Table 5-4

Limit Table Entries for Insertion Loss

Segment

Number

Limit

Type

Beginning

Stimulus

Ending

Stimulus

Beginning

Response

Ending

Response

1

Min

300 KHz

1.3 GHz

−1.8 dB

−1.8 dB

2

Min

1.3 GHz

3 GHz

−2.3 dB

−2.3 dB

3

Min

3 GHz

6 GHz

−2.8 dB

−2.8 dB

4

Min

6 GHz

9 GHz

−3.8 dB

−3.8 dB