background image

5 -10

Agilent Technologies Z5623A Option H48

Performance Verification

Setting Up Limit Testing

Step 4. Type “Insertion Loss” in the Entry field as shown in

Figure 5-4

. You can

do this with the keyboard shown in the Title Entry window if you want,
but an easier way is to use the USB keyboard shown in

Figure 2-9 on

page 2-15

. After typing “Insertion Loss”, press the Enter key on the

USB keyboard (or the OK button in the Title Entry window) to close the
window and display the new title.

Figure 5-4

Title Entry Window With “Insertion Loss” Entered

Step 5. Click Scale on the menu bar. When the Scale menu appears, scroll

down to the Scale... button and click it. See

Figure 5-5

.

Figure 5-5

The “Scale...” Button