Disaster Recovery

Disaster Recovery

This section provides recovery procedures for the following problems:

Malfunctioning Command Window

Malfunctioning Network Analyzer

Malfunctioning Command Window

Symptoms

Any of the following symptoms indicates a malfunctioning command
window.

No ud0: prompt is returned when the ibclr command is executed.

An error message is returned when the ibdev command is
properly executed.

Upon turning on the network analyzer, the following message
appears: “The device at ‘AT-GPIB/TNT’ cannot be stopped because
a program is still accessing it.”

When the Recovery Procedure described below is performed, the
“AT-GPIB/TNT” icon observed in Step 6 has a red slash through it
(indicating that the device is presently disabled). However, when
Step 7 is performed, the pop-up menu has “Disable” as a menu
item (indicating that the device is presently enabled!).

Recovery Procedure

Step 1. Close every application that is running.

a. If an application’s window is open, click on the X in the upper

right-hand corner of the window.

b. If an application’s window has been minimized, locate its icon on the

task bar at the bottom of the network analyzer screen. See

Figure

3-8

. Click on the icon. This will open the application’s window. Next,

click on the X in the upper right-hand corner of the window to close
the application.

c. When all applications are closed, the task bar will look like the one

in

Figure 3-9

.