Agilent Technologies Option H48 Multiport Test Set Z5623A User Manual

Page 61

Advertising
background image

Agilent Technologies Z5623A Option H48

4-11

Calibrating the Network Analyzer

Renaming the Calibration File

Step 9. Move the cursor over the filename that was copied down in

Step 16

of

Performing the Calibration

” on

page 4-6

. (If you were not able to copy

down the filename, select the “.cst” file whose date and time correspond
to when you completed the calibration.) Do not click the filename you
have selected. Just keep the cursor positioned over the filename for
about two seconds (until it is highlighted as shown in

Figure 4-18

).

Figure 4-18

Cal File Highlighted

Step 10. Move the cursor off the selected filename (either to the left or to the

right). The selected filename will remain highlighted.