Agilent Technologies Option H48 Multiport Test Set Z5623A User Manual

Page 87

Advertising
background image

Agilent Technologies Z5623A Option H48

5 -19

Performance Verification

Setting Up Limit Testing

Limit Testing for Return Loss (Port Active)

Recalling the Calibration

Recall the calibration file listed in

Table 4-1 on page 4-13

. (Consult

How to Recall a Calibration

” on

page 4-14

.)

Specifying Measurement Type, Title, and Scaling

The desired measurement type for the Return Loss test is S

11

. The

network analyzer is already configured to measure S

11

.

Step 1. Click View on the menu bar. When the View menu appears, scroll down

to the Meas Display button and pause momentarily. When the Meas
Display menu appears, scroll down to the Title... button and click it.
See

Figure 5-2

.

Step 2. The Title Entry window is shown in

Figure 5-3

. When this window pops

up, use the mouse to click anywhere in the white rectangle to the right
of the word, “Entry:”. (The white rectangle is referred to as the Entry
field.) A text entry cursor (a vertical line) will then appear in the white
rectangle.

Step 3. Type “Return Loss (Port Active)” in the Entry field as shown in

Figure

5-19

. You can do this with the keyboard shown in the Title Entry

window if you want, but an easier way is to use the USB keyboard
shown in

Figure 2-9 on page 2-15

. After typing “Return Loss (Port

Active)”, press the Enter key on the USB keyboard (or the OK button in
the Title Entry window) to close the window and display the new title.

Figure 5-19

Title Entry Window With “Return Loss (Port Active)” Entered

Step 4. Click Scale on the menu bar. When the Scale menu appears, scroll

down to the Scale... button and click it. See

Figure 5-5

.