Agilent Technologies Option H48 Multiport Test Set Z5623A User Manual

Page 89

Advertising
background image

Agilent Technologies Z5623A Option H48

5 -21

Performance Verification

Setting Up Limit Testing

Figure 5-21

Completed Limit Table for Return Loss (Port Active)

Turning On Limit Lines and Limit Testing

Step 1. Click Trace on the menu bar. When the Trace menu appears, scroll

down to the Limit Test... button and click it as shown in

Figure 5-8

.

The Limit Test window shown in

Figure 5-9

will pop up.

Step 2. Click the Limit Test ON and Limit Line ON boxes. The result is

shown in

Figure 5-12

.

Step 3. Click the Hide Table button, then click the OK button. The result is

shown in

Figure 5-22

. The trace shown in the figure is an S

11

measurement of the Electronic Calibration (ECal) module. Notice that
the ECal module does not meet the return loss specifications of the test
set.

Figure 5-22

Return Loss (Port Active) Limit Test On ECal Module