3-8

Agilent Technologies Z5623A Option H48

Using the Network Analyzer to Control the Test Set

Common Errors in Manual Commands

NOTE

It is possible to issue several commands in a single line. For example,
the following string connects the Reflection port to Test Port #2,
connects the Transmission port to Test Port #5, and sets the Control
Lines to OLOOL: “refl_02;tran_05;00001001;”<Enter>.

Common Errors in Manual Commands

This section explains several common errors.

Table 3-1

Common Errors in Manual Commands

Command(s) Issued

Explanation of Error

Solution

“refl_01”

Missing semicolon.

Send a semicolon:

“;”

“refl_01”

“tran_08;”

Missing semicolon in first
command causes both
commands to be interpreted
as a single invalid
command: “refl_01tran_08;”

Resend both
commands with a
semicolon in each:

“refl_01;”

“tran_08;”

“10101;”

Three leading zeroes are
missing. (This is a binary
Control Line command. It
needs three leading zeroes.)
(Decimal Control Line
commands don’t need
leading zeroes.)

Resend the command
with leading zeroes
included:

“00010101;”