Agilent Technologies Z5623A Option H48

3-5

Using the Network Analyzer to Control the Test Set

Preparing the Analyzer to Control the Test Set

Step 9. When the

C:\Progam Files\National Instruments\ NI-488.2\

bin\ibic.exe

window appears, type

ibdev

. See

Figure 3-6, “User

Controller Window.”

. Then you will be prompted to:

Enter board index: “0”

Enter primary address: “12”

Enter secondary address: “0”

Enter timeout: “0”

Enter ‘EOI on last byte’ flag: “0”

Enter end-of-string byte/mode: “1”

Figure 3-6

User Controller Window

Step 10. If you make a mistake, close the window and repeat steps 7, 8, and 9.

Step 11. When the prompt

ud0

:

appears in the window, enter

ibclr

. This

command clears the internal bus. When the ud0: prompt is returned
after the ibclr command has been issued, you are ready to issue
commands to the test set from this window.