Agilent Technologies Option H48 Multiport Test Set Z5623A User Manual
Page 79
Advertising
Agilent Technologies Z5623A Option H48
5 -11
Performance Verification
Setting Up Limit Testing
Step 6. The Scale window shown in
Figure 5-6
will pop up. When it does, set
the following values:
• Scale Per Division = 1 dB
• Reference Level = 0 dB
• Reference Position = 9 Divisions
When these values have been entered, the Scale window will look like
Figure 5-7
. Click the OK button.
Figure 5-6
The Scale Window With Default Values
Figure 5-7
The Scale Window With Desired Values For Insertion Loss