5 -14

Agilent Technologies Z5623A Option H48

Performance Verification

Setting Up Limit Testing

Turning On Limit Lines and Limit Testing

Step 1. Click Trace on the menu bar. When the Trace menu appears, scroll

down to the Limit Test... button and click it as shown in

Figure 5-8

.

The Limit Test window shown in

Figure 5-9

will pop up.

Step 2. Click the Limit Test ON and Limit Line ON boxes. The result is

shown in

Figure 5-12

.

Figure 5-12

Turning On Limit Testing and Limit Line Display

Step 3. Click the Hide Table button, then click the OK button. The result is

shown in

Figure 5-13

. The trace shown in the figure is an S

21

measurement of the Electronic Calibration (ECal) module. Notice that
the ECal module does not meet the insertion loss specifications of the
test set.