Agilent Technologies Option H48 Multiport Test Set Z5623A User Manual

Page 117

Advertising
background image

Agilent Technologies Z5623A Option H48

7 -5

Advanced Topics

Using an External Computer to Control the Test Set

Using National Instruments VISA

If you are using National Instruments VISA, be sure to set the following
variables as follows:

VI_ATTR_SEND_END = VI_FALSE ‘

This specifies whether to assert END

during the transfer of the last byte of the buffer

VI_ATTR_TERMCHAR = 0x0A ‘

This is the termination character. When

the termination character is read and

VI_ATTR_TERMCHAR_EN

is enabled

during a read operation, the read operation terminates.

VI_ATTR_TERMCHAR_EN = VI_TRUE ‘

This is a flag that determines

whether the read operation should terminate when a termination
character is received.

VI_ATTR_SUPPRESS_END_EN = VI_FALSE ‘

Specifies whether to

suppress the

END

bit termination. If this attribute is set to

VI_TRUE

, the

END

bit does not terminate read operations. If this attribute is set to

VI_FALSE

, the

END

bit terminates read operations.

Write Commands:

Append all commands with “\n,” for example,

*rst\n

.

Read Commands:

The Z5623A Option H48 returns data terminated by

\r\n

.