Agilent Technologies Option H48 Multiport Test Set Z5623A User Manual

Page 105

Advertising
background image

Agilent Technologies Z5623A Option H48

5 -37

Performance Verification

Verifying Return Loss and Insertion Loss Specifications

Step 7. Command the test set to terminate all of its test ports.

Step 8. Recall the limit test file for return loss (port off). It is listed in

Table 5-3

on page 5-7

.

Step 9. Read the results of the return loss (port off) test on the screen of the

network analyzer. Record the results in

Table 5-8

.

Step 10. Move the RF cable to the next test port. Repeat

Step 2

through

Step 9

.

Step 11. Repeat

Step 10

until all eight test ports have been measured.

Measurements Using The Transmission Port

Step 1. Move the RF cable from the Reflection Port to the Transmission Port.

Move the test port cable back to Test Port 1.

Step 2. Command the test set to connect the test port to the Transmission Port.

Consult Table 3-2 on page 3-9 for a list of commands.

Step 3. Recall the limit test file for insertion loss. It is listed in

Table 5-3 on

page 5-7

.

Step 4. Read the results of the insertion loss test on the screen of the network

analyzer. Record the results in

Table 5-8

.

Step 5. Recall the limit test file for return loss (port active). It is listed in

Table

5-3 on page 5-7

.

Step 6. Read the results of the return loss (port active) test on the screen of the

network analyzer. Record the results in

Table 5-8

.

Step 7. Move the RF cable to the next test port. Repeat

Step 2

through

Step 6

.

Step 8. Repeat

Step 7

until all eight test ports have been measured.