Electrical Requirements

Electrical Requirements

The alternating-current (AC) power that is supplied to the test set must
meet the following requirements:

Voltage: 90 to 250 Vac

Frequency: 48 to 66 Hz

Available power: 40 watts minimum

If the available AC line voltage is outside the 90 to 250 Vac range, an
autotransformer that provides third wire continuity to earth ground
may be used.

Environmental Requirements

Operating Environment

Indoor use only

Operating temperature: 0 to 55 °C

Maximum relative humidity: 80 percent for temperatures up to 31 °C
decreasing linearly to 50 percent relative humidity at 40 °C

Altitude: up to 15,000 feet (4,572 meters)

Enclosure protection: IP 20, according to IEC 529

CAUTION

This product is designed for use in INSTALLATION CATEGORY II,
and POLLUTION DEGREE 2, per IEC 101 and 664 respectively.

Non-Operating Storage Conditions

Temperature: –40 °C to +70 °C

Humidity: 0 to 90 percent relative at +65 °C (non-condensing)

Altitude: 0 to 15,240 meters (50,000 feet)