where (in the Source Memory) to store the calibration factors

calculated during the calibration task.

Fig.9.2 The state of the Source Memory after completion of a

CALIBRATION TASK

After completion of a calibration task the calibration factors, which

are found in the Working Memory, are entirely dependent upon the

Filter Bank No., which is active for the optical filter. If Filter Bank

No. 1 is active for the filter in positions “A” then calibration factors

from files A 1, W will be found in Working Memory (see Fig 9.2).

9.1.5

Effect of Changing the UA number in the “Active”

Set-up

WARNING!: Changing the UA number of the filter in the Monitor’s

active set-up will destroy all calibration data for the filter.

As explained in

section 9.1.1

, the UA No. determines the set of

Optical Filter Factors used during the calculation of calibration factors

and therefore, if the UA number of the filter is changed in the active

set-up of the Monitor, all calibration data related to the optical filter

has been removed from the set-up is automatically deleted from

both the Source Memory and the Working Memory as it is no

longer valid.