2 performing a span calibration – stand-alone use, Perform zero-point calibration? no yes, Connect – LumaSense Technologies 3434i SF6 Leak Detector User Manual

Page 165: And press enter when ready

Advertising
background image

Chapter 12

______________________________________________________________________

_____________________________________________________________________________

BE6028-14

3434i SF6 Leak Detector

LumaSense Technologies A/S

Page 165 of 191

1. For each filter being span calibrated, enter the No. of the filter

bank where you wish the Monitor to store the conversion

factors which will be calculated during the span calibration task.

2. Press MEASURE S3 S3 S1 then the following text will be

displayed:

PERFORM ZERO-POINT CALIBRATION?

NO YES

3.

If you only wish to span calibrate filters, follow the set-up of a

span calibration task by following the black arrows in the

schematic diagram shown in

Fig.12.15

.

The following text will appear on the display:

CONNECT

(the name of the gas appears here)

XXXXXXmg/m

3

AND PRESS ENTER WHEN READY

Note: if the gas concentration on the “Analysis Certificate” is given in

ppm then the following formula can be used to convert from parts

per million units (ppm) to mg/m

3

:

For a gas at 20°C and at 1 atmosphere pressure:

Concentration (in mg/m

3

) = Concentration (in ppm) x Molecular Weight (in g)

24,04

12.8.2 Performing a Span Calibration – Stand-alone Use

The general equipment necessary to perform a calibration task is

described in

section 12.4

. Follow the procedure from step 1 to step 3

inclusive and then continue as follows:

1.

a. Connect the free-end of the Teflon tubing mentioned in step 3

to the pressure valve on a cylinder of calibration gas.


b. Press

¿

.

c. Gently open the pressure-valve on the gas cylinder and

regulate the flow of gas so that when the Monitor’s pump is

Advertising