LumaSense Technologies 3434i SF6 Leak Detector User Manual

Page 168

Advertising
background image

Chapter 12

______________________________________________________________________

_____________________________________________________________________________

BE6028-14

3434i SF6 Leak Detector

LumaSense Technologies A/S

Page 168 of 191

12.9.1 Setting-up a Zero-point Calibration of the Water-

vapour Filter

If you wish to zero-point calibrate the water-vapour filter you must

zero-point calibrate all installed optical filters. The concentration

offset factors calculated during zero-point calibration of the filter

“A” will be stored in the filter banks (the active during calibration).

Check, therefore, that you make the relevant filter bank active for

the filter before calibrating it. This is discussed in more detail in

section 12.3.3.

The procedure for setting-up a zero-point calibration of the water-

vapour filter is as follows:

1. Press Measure S3 S3 S1 then the following text will be

displayed:

PERFORM ZERO-POINT CALIBRATION ?

NO YES

2. Follow the set-up of a zero-point calibration of the water-vapour

filter by following the black arrows in the schematic diagram

shown in

Fig.12.16

.

The following text will appear:

ZERO-POINT CALIB. CONNECT CLEAN, WET AIR

AND PRESS ENTER WHEN READY