1 measurement set-up – stand-alone use – LumaSense Technologies 3434i SF6 Leak Detector User Manual

Page 81

Advertising
background image

Chapter 7

______________________________________________________________________

_____________________________________________________________________________

BE6028-14

3434i SF6 Leak Detector

LumaSense Technologies A/S

Page 81 of 191

Limit: enables you to define the gas concentration that will trigger

the alarms. The acceptable values are shown in Table 7.7.

Select limits: this defines the concentration level when the alarm

is triggered. This parameter must be defined.

Table 7.7 The range of acceptable values and default value of the alarm

limit for the gases

Display Text

Acceptable Values

Default

Value

GAS A HIGH ALARM LIMIT 1

n

à

_____PPM

0-1000000ppm

Blank

GAS A HIGH ALARM LIMIT 1

n

à

_____mg/m3

Depends on the molecular

weight of the gas

Blank

n

à

= A to W

The Rolling Average option enables you to prevent the alarm

being triggered by sudden, small fluctuation in the gas

concentrations by using values, which are averaged over a defined

period of time. The averaging time is defined in minutes. The

acceptable values are: 1 to 999 mins, and the default value is 10

mins.

Action: this provides you with 3 ways to trigger the alarm. One or

more of the options can be selected simultaneously.

-Monitor Beep: when an alarm is triggered the monitor makes a

beeping noise.

Monitor Relay 1 & Monitor Relay 2: selecting these options

enable you to connect the alarms to a variety of either audio (for

example, a siren) and/or visual (for example a flashing light)

alarm systems.

Active Alarms: only the gases, which you have selected and

activated, are shown in this field.

7.2.1

Measurement Set-up – Stand-alone Use

The philosophy of the instructions is described in

Chapter 3

.

To enter the Set-Up Tree, press SET-UP and the monitor displays

the following screen text.

Advertising