LumaSense Technologies 3434i SF6 Leak Detector User Manual

Page 84

Advertising
background image

Chapter 7

______________________________________________________________________

_____________________________________________________________________________

BE6028-14

3434i SF6 Leak Detector

LumaSense Technologies A/S

Page 84 of 191

Environment branch

12. Press S3 to enter the Environment branch.

13. Use the select push-buttons to select the desired flushing

mode.

If you select Auto, go to step 16.

If you select Fixed Time, go to step 14.

14. Use the direction push-buttons to define the Chamber Flushing

Time.

15. Use the select push-buttons to select Tube Flushing.

If you don’t want tube flushing, select NO and go to step 17.

If you select tube flushing, use the direction push-buttons to

define the flushing time and go to step 17.

16. Use the direction push-buttons to select the desired sample

tube length and move on to the next display.

17. Use the direction push-buttons to select the desired

normalization temperature.

See

section 7.1

for details about the normalization options.

When step 17 is completed, the display returns to the top of the

MEASUREMENT branch.

Press p to go to the top of the Set-up Tree.

If you want measurement data presented as averaged concentration

values, then go to step 19. If you want your measurement data

presented as sample concentration values, then press SET-UP to

exit the set-up function.

Format branch

If you are starting here, press SET-UP:

18. Press S2 to enter the FORMAT branch.

19. Use the direction push-buttons to enter the correct averaging

period.

20. Use the select push-buttons to choose if you want to print each

gas.

If you select Yes, go to step 21.

If you select No, go to step 22.