Gas Monitor Display / Averaging Period (3434i stand-alone use):

this enables you to decide if instantaneous values or averaged values

are displayed on the gas monitors screen.

If average values are desired, then the example below illustrates

how the values are calculated and displayed.

Let us suppose that the Monitor has been taking measurements

every 2.5 min., over a period of 30 min., of gas A Sulphur

hexafluoride and the following results (un-shaded area) were

obtained and recorded in the display memory, for stand-alone

users.

A: Sulphur hexafluoride

Averaged Values

0

11:42:30

1

66.72 mg/m

3

66.72

mg/m

3

0

11:45:00

2

154.90 mg/m

3

110.9

mg/m

3

0

11:47:30

3

271.30 mg/m

3

164.3

mg/m

3

0

11:50:00

4

415.90 mg/m

3

227.2

mg/m

3

0

11:52:30

5

588.70 mg/m

3

299.5

mg/m

3

0

11:55:00

6

789.70 mg/m

3

444.1

mg/m

3

0

11:57:30

7

1.019

g/m

3

616.9

mg/m

3

0

12:00:00

8

1.276

g/m

3

817.9

mg/m

3

0

12:02:30

9

1.562

g/m

3

1.047

g/m

3

0

12:05:00

10

1.876

g/m

3

1.304

g/m

3

0

12:07:30

11

2.218

g/m

3

1.159

g/m

3

0

12:10:00

12

2.589

g/m

3

1.904

g/m

3

Let us suppose that the averaging period was set to 10 min. When

the above measurement results are displayed on the screen and

the averaging function is activated then each measurement will be

averaged over the previous 10 min. and the averaged

measurement results will replace the individual measurement

results (averaged results are shown in the shaded area in the list

above). For example: measurement number 7 would be 616.9

mg/m

3

. This is the average of the 5 measurements taken in the

10 min. period ending at 11:57:30. Measurement number 12

would be 1.904 g/m

3

which is the average of the 5 measurements

taken in the 10 min. period ending at 12:12:00.

The range of acceptable values and the default value of the

averaging period are shown in

Table 7.6

.

Advertising