Chapter 5

______________________________________________________________________

_____________________________________________________________________________

BE6028-14

3434i SF6 Leak Detector

LumaSense Technologies A/S

Page 68 of 191

and/or visual (for example a flashing light) alarm systems and

they are activated whenever a pre-defined concentration (alarm

level) of one (or more) of the gases being measured is exceeded.

Alarm signals are given by opening and closing the relay contacts.

Both relays are closed when the Monitor is switched off and when

the Monitor is switched on. This means that there is a direct

electrical connection between pins 1 & 2 and pins 4 & 5 at all

times except when the Monitor measures a gas concentration

which is greater than the corresponding alarm limit for the gas.

When this concentration is measured, the Monitor breaks the

connection between pins 1 & 2 or 4 & 5. The monitor will

furthermore activate the internal beeper and write the following

message on the display:

The user can switch off the beeper and the message by pressing

INFO or RESET push-button on the front-panel of the Monitor.

The alarm relays will continue being activated until either (1) all

measured gas concentrations fall below their user-defined alarm

level(s); or (2) the monitoring task is stopped – this automatically

closes the alarm relays.

Alarm Relay 1:

Alarm Relay 1: correspond to High Alarm Limit 1.

Whenever one or more gases exceed their alarm level(s) for high

alarm limit 1, the Monitor breaks the electrical connection

between pins 1 & 2 and this activates the attached alarm system.

Alarm Relay 2:

Alarm Relay 2: correspond to High Alarm Limit 2.

Whenever one or more gases exceed their alarm level(s) for high

alarm limit 2, the Monitor breaks the electrical connection between

pins 4 & 5 and this activates the attached alarm system.

LumaSense supplies a 6-pin DIN plug (male) with a locking collar

JP0600 for connecting external alarm devices to the alarm relay.

Caution:

The DC voltage across the relay contacts must not exceed 25V. The

potential on the relay contacts must not be more than 25VDC above

chassis potential, as this will cause an excessive leakage current. The

WARNING:

ALARM LIMIT EXCEEDED