Chapter 4

______________________________________________________________________

_____________________________________________________________________________

BE6028-14

3434i SF6 Leak Detector

LumaSense Technologies A/S

Page 32 of 191

cycle to another due to the cleanliness of the filters and the degree

of fluctuation in the concentration of the various gases.

Table 4.1 Approximate measurement times

Monitor Setup

Volume of Air

Response Times

S.I.T.: “Normal” (5s)

SIT : “Low Noise”

(20s)

Flushing:

Auto, (tube 1m)

140 cm

3

/sample

One gas: ~27s

One gas +water

~40s

One gas +water

~70s

S.I.T.: “Fast” (1s)

Flushing:

Chamber 4s, Tube

“OFF”

20cm

3

/sample

One gas: ~13s

One gas + water

~22s

4.1.2

Monitoring Task

A monitoring task consists of a series of similar measurement cycles.

These cycles can either be repeated at regular intervals or be

continuous:

4.1.3

Sampling Interval

The sampling interval is a measure of the time between the start of

one measurement cycle and the start of the following measurement

cycle.

When the Monitor is not sampling continuously the sampling interval

has to be defined by the user and is a fixed period of time (see

Fig.4.1

).