Section 12.11 – LumaSense Technologies 3434i SF6 Leak Detector User Manual

Page 173

Advertising
background image

Chapter 12

______________________________________________________________________

_____________________________________________________________________________

BE6028-14

3434i SF6 Leak Detector

LumaSense Technologies A/S

Page 173 of 191

12.9.4 Performing a Span Calibration of the Water-vapour

Filter

Section 12.4.1

describes a good method of producing water-vapour

of known concentration. The directions for performing a span

calibration of water-vapour filter are similar to those for performing a

humidity-interference calibration of the other installed filters is as

described in

section 12.7.2

. In both calibration tasks the signal in the

cell is measured when a supply of zero gas containing a constant

concentration of water vapour is attached to the Monitor. There are,

however, two differences:

1. During a humidity-interference calibration the signal is measured

using the installed filter (including the water-vapour filter),

whereas during the span calibration of the water-vapour filter the

signal is only measured with the water-vapour filter.

2. During a humidity-interference calibration, it is not necessary to

know the concentration of water-vapour supplied to the cell as it

is only necessary to measure the size of each signal (with filter

“A”) relative to the size of the signal measured with the water-

vapour filter. However, during a span calibration of the water-

vapour filter it is necessary to know the exact concentration of

water-vapour supplied to the cell, because during this calibration

the conversion factor is calculated. This factor relates the

measured signal to the concentration of water-vapour in the cell.

Follow the procedure described in

section 12.7.2

.

12.10 Setting-up a Combined Zero-point and

Humidity-interference Calibration – Stand-

alone Use

We have explained in previous sections of this chapter (

section 12.6

and

section 12.7.1

) that a zero-point calibration can only be

performed alone if a humidity gain factor is already stored in the

filter bank which are active during the zero-point calibration task.

Likewise, a humidity-interference calibration can only be performed

alone if a concentration offset factor is already stored in the filter

bank which is active during the humidity-interference calibration

task.

If the necessary calibration factors are not stored in the Monitor’s

memory, a zero-point calibration task has to be performed together

with a humidity-interference calibration task.

Advertising