LumaSense Technologies 3434i SF6 Leak Detector User Manual

Page 139

Advertising
background image

Chapter 12

______________________________________________________________________

_____________________________________________________________________________

BE6028-14

3434i SF6 Leak Detector

LumaSense Technologies A/S

Page 139 of 191

calibration factors compare with the “new” calibration factors

calculated during re-calibration tasks, see

section 12.3.1

.

Fig.12.3

illustrates only part of a typical calibration data print-out.

Note that there are five different filter banks, which contain

calibration data. These banks are numbered from 1 to 5. This

enables each filter to be calibrated to measure up to 5 different

gases. Before the filter is calibrated to measure any one particular

gas, you have to inform the Monitor about where you wish to store

the calibration data for this gas. The number of the filter bank chosen

when operating in Set-Up mode is called the active filter bank (see

Chapter 9

).

When checking the optical filter using the front panel push-buttons,

Chapter 10

provides details of how to obtain a print-out of calibration

data using a printer with an RS 232 interface.

Note that calibration data in the active filter bank is always printed

out directly after the optical filter factors data and this data is also

shown under the heading CALIBRATION DATA IN FILTER BANK

: X, where X is the number of the filter bank.

Note: all calibration factors are expressed as exponential numbers.

For example: 104.05E-3, this is the same as the number 104.05 x

10-3 = 0.10405

1. Refer to your calibration data print-out and check the following

information:

a. Under

the

heading

GENERAL

CALIBRATION

INFORMATION FOR FILTER A: check that the UA number

of the installed optical filter is UA0988 the same as the UA

number appearing on the “Calibration Chart” for the optical

filter in position “A”.

Under the headings CALIBRATION DATA IN FILTER

BANK:1; CALIBRATION DATA IN FILTER BANK:2;

CALIBRATION DATA IN FILTER BANK:3; CALIBRATION

DATA IN FILTER BANK:4; and CALIBRATION DATA IN

FILTER BANK:5 check the following information:

b. Zero-point calibration performed: if a date appears here

(year-month-day) this is the last time the zero-point

calibration data in this filter bank was updated.

c.

Concentration offset factor: if a number appears here this

means that this filter bank contains calibration data obtained

during a zero-point calibration of filter “A”.

d. Hum. Interference Calibration performed: if a date

appears here (year-month-day) this is the last time the

Advertising