Chapter 8

______________________________________________________________________

_____________________________________________________________________________

BE6028-14

3434i SF6 Leak Detector

LumaSense Technologies A/S

Page 94 of 191

u

&t

Using these push-buttons users can scroll “horizontally”

through any vertical level of data. For example, if the statistical

analysis of the measurement results for water-vapour are displayed

on the screen, then by using u you can read the statistical analysis

of the measurement results for gas A,W and then go back to the

original display (see

Fig.8.1

).

EVENT

This push-button can be used (pressed) to mark

the measurement being performed at the time at

which a special event occurred during a monitoring

task. Any such measurement will appear on the

display with an asterisk alongside it. By pressing

INFO when such a display is seen you can display

what the asterisk indicates.

Section 8.2.2

shows

the display which appears when INFO is pressed

and explains all the symbols used. Marked events

are also shown on print-outs and of monitoring

task data (see

section 8.2.3

).

q

This push-button allows the user to scroll

“vertically” downwards through the list of

measurement results (see

Fig.8.1

).

p

This push-button allows the user to scroll

“vertically” upwards through the list of

measurement results. The measurement list is a

closed loop.

The push-buttons p and q have a repetitive function, which means

that if they are pressed continuously they will allow the user to scroll

very quickly through a list.

8.1.1

Processing Measurement Data

Gas measurement data stored in Display Memory can be processed

in the following way:

·

It can be processed to produce Time-weighted Average (TWA)

concentrations of gases over any user-defined period of time.

8.1.2

Obtaining Time-weighted Average (TWA) Values

By using the AVERAGE button Time-weighted Average

concentrations (C

TWA

) of all the monitored gases can be obtained

over any user-defined period of time (T hours). The following steps

are involved:

1. Press SET-UP and S2.