1 retaining existing calibration factors – LumaSense Technologies 3434i SF6 Leak Detector User Manual

Page 148

Advertising
background image

Chapter 12

______________________________________________________________________

_____________________________________________________________________________

BE6028-14

3434i SF6 Leak Detector

LumaSense Technologies A/S

Page 148 of 191

calibration set-up. The calibration tasks are set-up while operating in

Measure mode.

Notice the following:

1. Before starting to set-up a calibration task you must make sure

that you choose the correct FILTER BANK NO. to store the

calibration factors which will be calculated during calibration (see

section 12.3.3).

2. If you answer “YES” to PERFORM ZERO-POINT CALIBRATION

?, then the installed optical filters (“A”) will be zero-point

calibrated, and the concentration offset factor for the filter will

be stored in the filter bank no. which is active for that filter

during the practical calibration task.

3. If you answer “YES” to PERFORM HUM. INTERFERENCE

CALIBRATION ?, then the installed optical filters (“A”) will be

calibrated for humidity interference, and the humidity gain

factor for the filter will be stored in the filter bank no. which is

active for that filter during the practical calibration task.

4. If you answer “YES” to PERFORM GAS SPAN CALIBRATION ?,

then you can choose (1) whether this calibration should be a

single-point or a two-point span calibration. The concentration

conversion factor calculated during span calibration for a par-

ticular gas using a particular filter), is stored in the filter bank no.

which is active during the calibration procedure.

12.5.1 Retaining Existing Calibration Factors

In some case, you may wish to keep the calibration factors already

present in your monitor. If this is the case, use the following

instructions.

Advertising