LumaSense Technologies 3434i SF6 Leak Detector User Manual

Page 150

Advertising
background image

Chapter 12

______________________________________________________________________

_____________________________________________________________________________

BE6028-14

3434i SF6 Leak Detector

LumaSense Technologies A/S

Page 150 of 191

“A”,is installed in your Monitor and have been calibrated. Suppose

the calibration data block of your Monitor contains the information

shown in Fig.12.8 (notice no filter is installed in position “E”)

Suppose You now install an optical filter in position “E”. This filter has

to be fully calibrated . Before performing a zero-point and humidity

interference task one of the following filter banks must be made

active:

·

Filter bank no. 3, 4 or 5 for filter “A”;

·

Filter bank no. 2, 3 or 5 for filter “B”;

·

Filter bank no. 2, 3 or 4 for filter “C”;

·

Filter bank no. 2, 3, 4 or 5 for filter “D”.

Suppose you make filter bank no. 3 active for all the above filters

and filter bank no. 1 active for filter “E”. Then, after a zero-point

calibration the calibration data block would contain the information

shown in

Fig.12.9

. None of the original calibration data has been

overwritten by the newly-calculated calibration factors.

Note: 3434i only have an optical filter in position A

Fig.12.8 Illustration of the calibration data stored in the filter banks

of filters “A” to “D” before a zero-point calibration task